Chapter 3 Specifications
Introduction
Operating Guide
3-17
General Information
Save and Recall:
Up to 20 complete instrument setups may be saved and recalled later. These setups are retained when
power is removed from the counter.
Rack Dimensions (HxWxD):
88.5 mm x 212.6 mm x 348.3 mm
Weight:
3.5 kg maximum
AC Line Supply
DC Supply (Option 002 Only)
Power Supply Voltage:
100 to 120 VAC ±10% - 50, 60 or 400 Hz ±10%
10 to 32 VDC, 3-pin male XLR connector
220 to 240 VAC ±10% - 50 or 60 Hz ±10%
AC Line Voltage Selection: Automatic
Option 002 may not be ordered with Option 060
Power Requirements:
170 VA maximum (30 W typical)
4A initial inrush at 10 VDC
3A max, once stabilized
Operating Environment:
0° C to 55° C
Storage Environment:
–40° C to 71° C
Remote Interface:
GPIB (IEEE 488.1-1987, IEEE 488.2-1987)
GPIB Interface Capabilities:
SH1, AH1, T5, TE0, L4, LE0, SR1, RL1, PP0, DC1, DT1, C0, E2
Remote Programming Language: SCPI-1992.0 (Standard Commands for Programmable Instruments)
Safety:
Designed in compliance with IEC 1010-1, UL 3111-1 (draft), CAN/CSA 1010.1
EMC:
CISPR-11, EN50082-1, IEC 801-2, -3, -4
Electrostatic Discharge and Fast Transient/Burst Immunity Testing: When the product is operated at
maximum sensitivity (20 mVrms) and tested with 8kV AD according to IEC801-2 or with 1kV power line
transients according to IEC 801-4, frequency miscounts may occur that will affect measurement data
made during these disturbances.
Radiated Immunity Testing: When the product is operated at maximum sensitivity (20 mVrms) and tested
at 3 V/m according to IEC 801-3, external 100 to 200 MHz electric fields may cause frequency miscounts.
RS-232C:
The rear-panel RS-232 connector is a 9-pin connector (DB-9, male). You can connect the universal
counter to any terminal or printer with a properly configured DTE connector (DB-25). You can use a
standard interface cable (Agilent part number 24542G or 24542H). Data is “output only”; the instrument
can not be programmed via the RS-232 interface.
Note on Pin 4:
May be used as either a DTR signal or an indication of measurement in-limit as configured by the Utility menu.
When used as an in-limit indicator, the signal will be high for every measurement within the user set limits.
Pin Number
Type
Description
2
Input
Receive Data (RxD) (for Xon/Xoff only)
3
Output
Transmit Data (TxD)
4
Output
Data Terminal Ready (DTR)
Measurement In-Limit Signal
5
—
Signal Ground
6
Input
Data Set Ready (DSR)
*
All other pins: no connection
1 2 3 4 5
6 7 8 9
Summary of Contents for 53181A
Page 2: ...NOTES...
Page 14: ...Contents x Operating Guide...
Page 24: ...Preface xx Operating Guide...
Page 33: ...1 Getting Started...
Page 68: ...Chapter 1 Getting Started Making Measurements 1 36 Operating Guide 1...
Page 69: ...2 Operating Your Frequency Counter Operator s Reference...
Page 143: ...3 Specifications...
Page 160: ...Chapter 3 Specifications Introduction 3 18 Operating Guide...