Chapter 1 Getting Started
Making Measurements
1-28
Operating Guide
1
Setting the Counter to Flag On Limits But Continue Measuring
Perform the following steps to select the GO ON choice in the ON FAIL display if
you want the Counter to continue measuring even though an measurement result
exceeds the limits previously entered.
The current modified measurement of the input signal applied to CHANNEL 1 is
displayed.
Since the Counter is now in the go-on-fail mode, the Limit annunciator in the display
will light each time a measurement exceeds the limits you set. However, the Counter
will continue to make measurements.
Disabling Limit Testing
The Counter is now making measurements without limit testing.
Limit
Modes
Run
Freq
Ch 1
Gate
M
Limit
Modes
Summary of Contents for 53181A
Page 2: ...NOTES...
Page 14: ...Contents x Operating Guide...
Page 24: ...Preface xx Operating Guide...
Page 33: ...1 Getting Started...
Page 68: ...Chapter 1 Getting Started Making Measurements 1 36 Operating Guide 1...
Page 69: ...2 Operating Your Frequency Counter Operator s Reference...
Page 143: ...3 Specifications...
Page 160: ...Chapter 3 Specifications Introduction 3 18 Operating Guide...