Chapter 2 Operating Your Frequency Counter
Using the Gate & External Arm Menu Key
Operating Guide
2-19
2
Example Procedure for Changing the Number of Digits of Resolution
Displayed for More Precise Measurements
1
Turn POWER key OFF then ON again (cycle POWER key) to preset the Counter.
2
Connect (for demonstration purposes) the 10 MHz Out signal to the channel 1
input.
3
Press Other Meas key until
PERIOD 1
is displayed.
After a few seconds,
0.100,000,000,0
µ
s
is displayed. Hence, the default number of
digits (or resolution) displays a precise measurement.
4
To demonstrate how to use “digits arming” to change the number of digits
displayed, perform the following steps first.
a.
Press Gate & ExtArm key
GATE: TIME
is displayed.
b.
Press any one of the arrow keys until
AUTO
is displayed.
c.
Press Enter or Run key.
0.100,0
µ
s
is displayed. Note that fewer digits are displayed.
In “auto arming” you produce results quickly. The short gate time produces a
lower-resolution result.