3- 46
Triggering the Oscilloscope
To use TV triggering
Triggering on Line 71
Line Numbers per Field for Each TV Standard
TV Standard
Field 1
Field 2
Alt Fld
NTSC
1 to 263
1 to 262
1 to 262
PAL
1 to 313
314 to 625
1 to 313
PAL-M
1 to 263
264 to 525
1 to 263
SECAM
1 to 313
314 to 625
1 to 313
GENERIC
1 to 1024
1 to 1024
1 to 1024 (Vertical)
Line Number Represents Count
In
Generic
mode, the line number represents the number of a count instead of a real
line number. This is reflected in the label in the softkey changing from
Line
to
Count
.
In the
Mode
softkey selections,
Line:Field 1
,
Line:Field 2
and
Count:Vertical
are used
to indicate where the counting starts. For an interlaced TV system, the counting
starts from the rising edge of the first vertical serration pulse of Field 1 and/or Field 2.
For a non-interlaced TV system, the counting starts after the rising edge of the
vertical sync pulse.