5- 40
Making Measurements
Setting measurement thresholds
4 Press the
Lower
softkey, then turn the Entry knob to set the lower
measurement threshold value.
Increasing the lower value beyond the set middle value will automatically
increase the middle value to be more than the lower value. The default lower
threshold is 10% or 800 mV.
If threshold
Type
is set to
%
, the lower threshold value can be set from 5% to 93%.
5 Press the
Middle
softkey, then turn the Entry knob to set the middle
measurement threshold value.
The middle value is bounded by the values set for lower and upper thresholds.
The default middle threshold is 50% or 1.20 V.
• If threshold
Type
is set to
%
, the middle threshold value can be set from 6%
to 94%.
6 Press the
Upper
softkey, then turn the Entry knob to set the upper
measurement threshold value.
Decreasing the upper value below the set middle value will automatically
decrease the middle value to be less than the upper value. The default upper
threshold is 90% or 1.50 V.
• If threshold
Type
is set to
%
, the upper threshold value can be set from 7%
to 95%.
Summary of Contents for 54621A
Page 4: ...iv ...
Page 10: ...Contents 6 ...
Page 11: ...1 Getting Started ...
Page 38: ...1 28 ...
Page 39: ...2 Front Panel Overview ...
Page 63: ...3 Triggering the Oscilloscope ...
Page 117: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Page 171: ...5 Making Measurements ...
Page 222: ...5 52 Making Measurements Making overshoot and preshoot measurements ...
Page 223: ...6 Utilities ...
Page 239: ...7 Performance Characteristics ...
Page 250: ...7 12 Performance Characteristics Agilent 54620 series Performance Characteristics ...
Page 260: ...7 22 ...