Performance Characteristics
Agilent 54620-series Performance Characteristics
FFT
Points
Fixed at 2048 points
Source of FFT
Analog channels 1 or 2 (or 3 or 4 for 54624A), 1+2, 1-2, 1*2
Window
Rectangular, Flattop, Hanning
Noise Floor
-70 to -100 dB depending on averaging
Amplitude Display
In dBV
Frequency Resolution:
0.097656/(time per div)
Maximum Frequency
102.4/(time per div)
Storage
Save/Recall (non-volatile)
3 setups and traces can be saved and recalled internally
Floppy Disk
Image formats
Data formats
Trace/setup formats
3.5” 1.44 MB double density
TIF, BMP
X and Y (time/voltage) values in CSV format
Recalled
I/O
RS-232 (serial) standard port
1 port; XON or DTR; 8 data bits; 1 stop bits; parity=none; 9600, 19200, 38400, 57600 baud
rates
Parallel standard port
Printer support
Printer Compatibility
HP DeskJet, HP LaserJet with HP PCL 3 or greater compatibility
Compatibility– black and white @150x150 dpi
gray scale @ 600x600 dpi
Epson– black and white @180x180 dpi
Seiko– DPU-414 black and white
Optional GPIB Interface Module Fully programmable with IEEE488.2 compliance
Typical GPIB throughput of 20 measurements or twenty 2000-point records per second.
* Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and
±10 ° C from firmware calibration temperature.
Summary of Contents for 54621A
Page 4: ...iv ...
Page 10: ...Contents 6 ...
Page 11: ...1 Getting Started ...
Page 38: ...1 28 ...
Page 39: ...2 Front Panel Overview ...
Page 63: ...3 Triggering the Oscilloscope ...
Page 117: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Page 171: ...5 Making Measurements ...
Page 222: ...5 52 Making Measurements Making overshoot and preshoot measurements ...
Page 223: ...6 Utilities ...
Page 239: ...7 Performance Characteristics ...
Page 250: ...7 12 Performance Characteristics Agilent 54620 series Performance Characteristics ...
Page 260: ...7 22 ...