7-13
Agilent 54640-series Performance
Characteristics
* Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and
±10 ° C from firmware calibration temperature.
Acquisition: Analog Channels
Max Sample rate
2 GSa/s interleaved, 1 GSa/s each channel
Max Memory Depth
8 MB interleaved, 4 MB each channel
Vertical Resolution
8 bits
Peak Detection
1 ns @ max sample rate
Averages
selectable from 2, 4, 8, 16, 32, 64 ...to 16383
High Resolution Mode
12 bits of resolution when
≥
100 us/div, average mode with average = 1
Filter:
Sinx/x interpolation (single shot BW = sample rate/4 or bandwidth of scope, whichever
is less) with vectors on.
Acquisition: Digital Channels (on 54641D and 54642D only)
Max Sample Rate
1 GSa/s
Max Memory Depth
4 MB
Vertical Resolution
1 bit
Glitch Detection
(min pulse width)
5 ns