3- 24
Triggering the Oscilloscope
To use I2C triggering
To use I
2
C triggering
An I
2
C (Inter-IC bus) trigger setup consists of connecting the oscilloscope to
the serial data (SDA) line and the serial clock (SCL) line, then triggering on a
stop/start condition, a restart, a missing acknowledge, an EEPROM data read,
or on a read/write frame with a specific device address and data value.
1 Press the
More
key in the Trigger section of the front panel, rotate the
Entry knob until
I
2
C
is displayed in the
Trigger
softkey, then press the
Settings
softkey to display the I
2
C trigger menu.
2 Connect an oscilloscope channel to the SCL (serial clock) line in the
circuit under test, then set the
SCL
clock channel softkey to that channel.
As you press the
SCL
softkey (or rotate the Entry knob on mixed-signal
oscilloscopes), the
SCL
label for the source channel is automatically set and the
channel you select is shown in the upper-right corner of the display next to
"
I
2
C
".
Adjust the trigger level for the selected analog channel by turning the Trigger
Level
knob. For the mixed-signal oscilloscope, select
Threshold
in the
D7 Thru
D0
or
D15 Thru D8
menu to set the level for digital channels assigned to the clock
and data lines. The value of the trigger level or digital threshold is displayed in
the upper-right corner of the display.
SCL Clock
channel
SDA Data
channel
Trigger on:
condition
Data
value
Trigger level
or threshold
Currently selected Clock
or Data channel
I
2
C trigger
Return to
previous menu
Address
Summary of Contents for 54621A
Page 4: ...iv ...
Page 10: ...Contents 6 ...
Page 11: ...1 Getting Started ...
Page 38: ...1 28 ...
Page 39: ...2 Front Panel Overview ...
Page 63: ...3 Triggering the Oscilloscope ...
Page 117: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Page 171: ...5 Making Measurements ...
Page 222: ...5 52 Making Measurements Making overshoot and preshoot measurements ...
Page 223: ...6 Utilities ...
Page 239: ...7 Performance Characteristics ...
Page 250: ...7 12 Performance Characteristics Agilent 54620 series Performance Characteristics ...
Page 260: ...7 22 ...