3- 26
Triggering the Oscilloscope
To use I2C triggering
EEPROM Data Read
: the trigger looks for EEPROM control byte value 1010xxx
on the SDA line, followed by a Read bit and an Ack bit. It then looks for the data
value and qualifier set by the
Data
softkey and the
Data is
softkey. When this
event occurs, the scope will trigger on the clock edge for the Ack bit after the
data byte. This data byte does not need to occur directly after the control byte.
Frame (Start: Addr7: Read: Ack: Data)
or
Frame (Start: Addr7: Write: Ack: Data)
: the oscilloscope triggers on a read or write
frame in 7-bit addressing mode on the 17th clock edge if all bits in the pattern
match. For triggering purposes, a restart is treated as a start condition.
SDA
SCL
Read
Ack
Ack
R/W
Data
Trigger point
Start or
Restart Condition
Control byte
SDA
SCL
Start or
Restart
Condition
Trigger point
17th clock edge
Address
Read
Write
R/W
Ack
Data
Ack
Stop
Condition