5- 7
Troubleshooting
To construct your own dummy load
To construct your own dummy load
1
Obtain a connector that is compatible with the connector J2A and J2B
on the Low Voltage Power Supply.
2
Connect load resistors to the connector as follows:
Figure 5-1
Location of the Low Voltage Power Supply Connectors
Connector
Load (Amps)
Resistor
Watts
Connect to Pin
Ground to Pin
+5.1 V
4.4 A
1.2
Ω
22 W
10, 11, 12
13, 14
-5.2 V
1.6 A
3.25
Ω
8 W
4, 5
6
+15.75 V
1.2 A
13.1
Ω
20 W
1
2
+3.4 V
4 A
0.825
Ω
13 W
15, 16, 17, 18
19, 20
-$
-
-%
5
-
E
Fan Connector
+3.4 V Adjustment
service.book Page 7 Wednesday, August 28, 2002 3:25 PM
Summary of Contents for 54641A
Page 4: ...iv service book Page iv Wednesday August 28 2002 3 25 PM ...
Page 7: ...1 General Information ...
Page 24: ...1 18 ...
Page 25: ...2 Preparing the Oscilloscope for Use service book Page 1 Wednesday August 28 2002 3 25 PM ...
Page 45: ...3 Testing Performance service book Page 1 Wednesday August 28 2002 3 25 PM ...
Page 66: ...3 22 service book Page 22 Wednesday August 28 2002 3 25 PM ...
Page 67: ...4 Calibrating and Adjusting service book Page 1 Wednesday August 28 2002 3 25 PM ...
Page 76: ...4 10 service book Page 10 Wednesday August 28 2002 3 25 PM ...
Page 77: ...5 Troubleshooting service book Page 1 Wednesday August 28 2002 3 25 PM ...
Page 90: ...5 14 service book Page 14 Wednesday August 28 2002 3 25 PM ...
Page 91: ...6 Replacing Assemblies service book Page 1 Wednesday August 28 2002 3 25 PM ...
Page 107: ...7 Replaceable Parts service book Page 1 Wednesday August 28 2002 3 25 PM ...
Page 114: ...7 8 service book Page 8 Wednesday August 28 2002 3 25 PM ...