5- 8
Troubleshooting
To check out the oscilloscope
To check out the oscilloscope
1
I
s an interface module connected to the rear panel of the oscilloscope?
If the answer is yes, follow these steps. Otherwise, go to step 2.
a
Turn off the oscilloscope.
b
Remove the interface module.
c
Turn on the oscilloscope, then check for the failing symptom.
d
If the failing symptom is not present, replace the interface module.
e
If the failing symptom is still present, go to step 2.
2
Disconnect any external cables from the front panel.
3
Disconnect the power cord, then remove the cabinet.
C A U T I O N
USE AN EXTERNAL FAN TO AVOID OVERHEATING COMPONENTS!
If you operate the oscilloscope with the cabinet removed, you must use an
external fan to blow air across the system board. This external air flow is
necessary to cool the heat sinks on the system board. Otherwise, damage to
the components can occur.
4
Connect the power cord, then turn on the oscilloscope.
• If the oscilloscope display comes on after a few seconds, you will need to
check the low voltage power supply. Go to “To check the Low Voltage Power
Supply” on page 5-11. Then return here.
• If, after checking the low voltage power supply, the voltages are within the
test limits, go to step 8. If they are not within the test limits, go to step 6.
If the display did not come on, follow these steps:
a
Disconnect the power cord.
b
Check to verify that all of the cable connections are secure.
c
Check the low voltage power supply. Go to “To check the Low Voltage
Power Supply” on page 5-11.
If the voltages are within the limits go to step 5. If the voltages are not within
the limits, go to step 6.
service.book Page 8 Wednesday, August 28, 2002 3:25 PM
Summary of Contents for 54641A
Page 4: ...iv service book Page iv Wednesday August 28 2002 3 25 PM ...
Page 7: ...1 General Information ...
Page 24: ...1 18 ...
Page 25: ...2 Preparing the Oscilloscope for Use service book Page 1 Wednesday August 28 2002 3 25 PM ...
Page 45: ...3 Testing Performance service book Page 1 Wednesday August 28 2002 3 25 PM ...
Page 66: ...3 22 service book Page 22 Wednesday August 28 2002 3 25 PM ...
Page 67: ...4 Calibrating and Adjusting service book Page 1 Wednesday August 28 2002 3 25 PM ...
Page 76: ...4 10 service book Page 10 Wednesday August 28 2002 3 25 PM ...
Page 77: ...5 Troubleshooting service book Page 1 Wednesday August 28 2002 3 25 PM ...
Page 90: ...5 14 service book Page 14 Wednesday August 28 2002 3 25 PM ...
Page 91: ...6 Replacing Assemblies service book Page 1 Wednesday August 28 2002 3 25 PM ...
Page 107: ...7 Replaceable Parts service book Page 1 Wednesday August 28 2002 3 25 PM ...
Page 114: ...7 8 service book Page 8 Wednesday August 28 2002 3 25 PM ...