5- 10
Troubleshooting
To check out the oscilloscope
8
Is the fan running?
The low voltage power supply has a thermal cut-out circuit. If the fan is
defective, the low voltage power supply will shut down when it gets too hot for
safe operation.
If the fan is running, perform the internal self-tests. Go to “To run the internal
self-tests” on page 5-13.
If the fan is not running, it may be defective. Follow these steps:
a
Disconnect the fan cable from the power supply.
b
Measure the fan voltage at the connector on the power supply.
See Figure 5-2 for the location of the fan connector.
c
If the fan voltage is approxi8 Vdc, replace the fan. If the fan voltage
is not approxi8 Vdc, replace the power supply.
Figure 5-2
Location of the Fan Connector
-$
-
-%
5
-
E
Fan Connector
+3.4 V Adjustment
service.book Page 10 Wednesday, August 28, 2002 3:25 PM
Summary of Contents for 54641A
Page 4: ...iv service book Page iv Wednesday August 28 2002 3 25 PM ...
Page 7: ...1 General Information ...
Page 24: ...1 18 ...
Page 25: ...2 Preparing the Oscilloscope for Use service book Page 1 Wednesday August 28 2002 3 25 PM ...
Page 45: ...3 Testing Performance service book Page 1 Wednesday August 28 2002 3 25 PM ...
Page 66: ...3 22 service book Page 22 Wednesday August 28 2002 3 25 PM ...
Page 67: ...4 Calibrating and Adjusting service book Page 1 Wednesday August 28 2002 3 25 PM ...
Page 76: ...4 10 service book Page 10 Wednesday August 28 2002 3 25 PM ...
Page 77: ...5 Troubleshooting service book Page 1 Wednesday August 28 2002 3 25 PM ...
Page 90: ...5 14 service book Page 14 Wednesday August 28 2002 3 25 PM ...
Page 91: ...6 Replacing Assemblies service book Page 1 Wednesday August 28 2002 3 25 PM ...
Page 107: ...7 Replaceable Parts service book Page 1 Wednesday August 28 2002 3 25 PM ...
Page 114: ...7 8 service book Page 8 Wednesday August 28 2002 3 25 PM ...