Calibration
Overview
Factory
Mainframe
Calibrations
The
temperature
change
is
displayed
at
the
top
of
the
display
as
shown
in
the
following
gure
.
Figure
3-1.
Current
Frame
1T
emp
condition
If
the
Current
Frame
1Temp
listing
is
greater
than
65
C,
then
the
mainframe
should
either
be
calibrated
at
the
current
operating
temperature
or
be
placed
in
an
ambient
air
temperature
that
is
within
5
C
of
the
temperature
of
the
current
calibration.
3-5