Maintenance
Agilent 55B AA Spectrometer User’s Guide
113
WARNING
Explosion Hazard
Failure to fill the liquid trap will cause violent and loud explosions to occur in
both the liquid trap and drain vessel when the flame is lit. The explosion may
be severe enough to cause death, personal injury or damage to the
instrument or laboratory.
NEVER attempt to light a flame unless the liquid trap is filled with solvent so
that the float just starts to move.
Figure 73
. Complete nebulizer/spray chamber assembly
The complete spray chamber/nebulizer assembly (refer Figure 73 above) is ready
to be installed in the sample compartment of the instrument.
Installing the spray chamber/nebulizer assembly
CAUTION
Always check the assembly before installing the spray chamber in the sample
compartment. Especially check the correct positioning of the mixing.