Operation
90
Agilent 55B AA Spectrometer User’s Guide
Output to LIMS system
Results can be sent to a LIMS system using the RS-232 output on the rear of the
Instrument.
To output to a LIMS system:
1
Connect a RS-232 cable from the rear of the instrument to the RS-232 input on
the LIMS, for example the COM1 connector on the rear of a PC.
2
Set the baud rate of the connection to that of the controlling system. This is
done at the Serial Port sub menu of the Options menu.
Each time the READ key is pressed data is output , the layout is as follows:
Calibration Standard ID, Concentration, Background Absorbance, Mean
Absorbance, Replicates Absorbances.
Batch Number, Sample Number, Concentration, Background Absorbance, Mean
Absorbance, Replicates Absorbances.
Shut down the system
Prior to shutting down the system at the end of the working day you should
aspirate pure solvent for 10 minutes.
To shut down the system:
1
Shut down any accessories according to their manuals.
2
Turn off the instrument.
3
Shut off all gas supplies at the regulators and gas cylinders.
4
Empty the waste vessel.
5
Turn off the exhaust fan.
6
Remove and Cleaning
7
Any hardware according to the instructions in the Maintenance Chapter – see
Page 93.