Alignment Procedures
101
Alignment Program (continued)
3130
PRINT "Calibrating output";Chan;"";I_range;"A current sink readback"
3140 !
3150
OUTPUT @Ps;"IRSET";Chan,I_range;";IRSET";Other chan,I_range
3160
! Set I range for source and
3170
! sink range
3180
OUTPUT @Ps; “ISET”;Other_chan,0;”;VSET”;Other_chan,7
3190
! Set source V & I output
3200
Output_on
3210
WAIT 1
! Wait for source to stabilize
3220
!
3230
OUTPUT @Ps;"IRLN";Chan
! Set sink readback low
3240
REPEAT
! Wait for supply to finish
3250
UNTIL BIT(SPOLL(@Ps),4)
! Finished when Bit 4 goes true
3260
Ilo=(FNDci(I_range))*Polarity
! Read low sink current
3270 !
3280
OUTPUT @Ps;"ISET";Other_chan,I_range
! Set source output current
3290
WAIT 1
3300
!
3310
OUTPUT @Ps;"IRHN";Chan
! Set sink readback high
3320
REPEAT
! Wait for supply to finish
3330
UNTIL BIT(SPOLL(@Ps),4)
! Finished when Bit 4 goes true
3340
Ihi=(FNDci(I_range))*Polarity
! Read high sink current
3350
!
3360
OUTPUT @Ps; NIDAT”;chan,Ilo,Ihi
! Send current sink data
3370
!
3380 SUBEND
3390 !
3400 !
3410 !
3420 Get_data SUB Get_data
! Subprogram with range data
3430
COM /Specs/ Vrng_lo,Vrng_hi,Irng_lo,Irng_hi
3440
COM /Ps/ Chan,Other_chan,@Ps,Model$[7]
3450
!
3460
IF Model$="Agilent6625A" AND Chan=1 THEN RESTORE Type_1_data
3470
IF Model$="Agilent6625A" AND Chan=2 THEN RESTORE Type_2_data
3480
IF Model$=”Agilent6626A" AND Chan=1 THEN RESTORE Type_1_data
3490
IF Model$="Agilent6626A" AND Chan=2 THEN RESTORE Type 1 data
3500
IF Model$=“Agilent6626A" AND Chan=3 THEN RESTORE Type_2_data
3510
IF Model$=“Agilent6626A" AND Chan=4 THEN RESTORE Type_2_data
3520
IF Model$=“Agilent6628A" OR Model$=“Agilent6629A" THEN RESTORE Type_2_data
3530
!
3540
READ Vrng_lo,Vrng_hi,Irng_lo,Irng_hi
! Read the Range data
3550
!
3560
!
3570
!
Output variables
range data
3580
!
Vrng Irng
3590
!
Lo Hi
Lo Hi
3600 Type 1 data:DATA
7, 50,
.015, .500
! Type 1 data (25W output)
3610 Type_2_data:DATA
16, 50,
.200, 2
! Type 2 data (50W output)
3620 !
3630 SUBEND
3640 !
Summary of Contents for 6625A
Page 10: ......
Page 24: ......
Page 30: ......
Page 44: ......
Page 62: ......
Page 68: ...Remote Operation 68 Figure 5 2 Sheet 1 of 2 Syntax Forms For Power Supply Commands ...
Page 69: ...Remote Operation 69 Figure 5 2 Sheet 2 of 2 Syntax Forms For Power Supply Commands ...
Page 90: ......
Page 94: ...Alignment Procedures 94 Figure A 3 Figure A 4 Figure A 5 ...
Page 116: ......
Page 120: ......