Error Codes and Messages
119
Table D-2. ERROR Responses (continued)
Error Code
(ERR? query)
Message
(ERR key)
Explanation
18
CAL LOCKED
Calibration was attempted with the calibration jumper on the GP-IB
board in the lockout position (See Section IV in the Service Manual).
Reposition the jumper and re-calibrate if this is desired.
22
SKIP SLF TST
The self test jumper on the GP-IB board is in the Skip Self Test Position
(See Section IV in the Service Manual). Reposition the jumper and carry
out self-test if this is desired.
Table D3. TEST? Response
Response Code
(TEST? query)
Explanation
0
No errors were detected.
20
The timer on the GP-IB board failed self test. Refer to the Troubleshooting Section in the
Service Manual.
21
The RAM on the GP-IB Board failed self test. Refer to the Troubleshooting Section in the
Service Manual. Note that if this failure occurred at power-on, it cannot be displayed on the
front panel.
27
The RAM on the GP-IB Board failed self test. Refer to the Troubleshooting Section in the
Service Manual. Note that if this failure occurred at power-on, it cannot be displayed on the
front panel.
29
See Table 5-9 on page 84.
30
See Table 5-9 on page 84.
Summary of Contents for 6625A
Page 10: ......
Page 24: ......
Page 30: ......
Page 44: ......
Page 62: ......
Page 68: ...Remote Operation 68 Figure 5 2 Sheet 1 of 2 Syntax Forms For Power Supply Commands ...
Page 69: ...Remote Operation 69 Figure 5 2 Sheet 2 of 2 Syntax Forms For Power Supply Commands ...
Page 90: ......
Page 94: ...Alignment Procedures 94 Figure A 3 Figure A 4 Figure A 5 ...
Page 116: ......
Page 120: ......