General Information
18
Table 1-3. Supplemental Characteristics (continued)
Output Power
25 Watt Output
50 Watt Output
Output Range
Lo Range
Hi Range
Lo Range
Hi Range
Output Volts
0-7 V
0-50 V
0-16 V
0-50 V
Output Amps
0-15 mA
0-500 mA
0-200 mA
0-2 A
Temperature Coefficient-Measurement:
Voltage
(0.002% + 0.1 mV)
per° C +0.5 mV
(0.002% + 0.1 mV)
per° C +4 mV
(0.002% + 0.1 mV)
per° C +1 mV
(0.002% + 0.1 mV)
per° C +4 mV
±
Current
(0.0025% + 1.5
µ
A)
per° C + 1
µ
A
(0.0025% + 10
µ
A)
per° C + 40
µ
A
(0.0025% + 20
µ
A)
per° C + 15
µ
A
(0.0025% + 50
µ
A)
per° C + 150
µ
A
Long Term Drift
: (In an 8 hour period following a 30 minute warm up)
Voltage
0.006% + 0.5 mV
0.006% + 2 mV
0.006% + 0.5 mV
0.006% + 2 mV
+ Current
0.01% + 5
µ
A
0.01% + 20
µ
A
0.01% + 40
µ
A
0.01% + 60
µ
A
Short Term Drift:
(Within 30 minutes after a line and/or load change):
Voltage
0.002 + 0.5 mV
0.002 + 1 mV
0.002 + 0.5 mV
0.002 + 1 mV
+ Current
0.01% + 2
µ
A
0.01% + 20
µ
A
0.01% + 25
µ
A
0.01% + 60
µ
A
DC Floating Voltage:
No output terminal may be more that 240 Vdc from any other terminal or from chassis ground. Also, no overvoltage
terminal may be more than 240 Vdc from any other terminal or chassis ground.
Remote Sense Capability:
(see pages 50 - 52):
Outputs can maintain specifications with up to 10 volt total in the remote sense leads.
AC Input Power
: 550 Watts
Programming Resolution:
Voltage
460
µ
V
3.2 mV
1 mV
3.2 mV
+ Current
1
µ
A
33
µ
A
13
µ
A
131
µ
A
OVP
230 mV
230 mV
230 mV
230 mV
Readback Resolution:
Voltage
483
µ
V
3.3 mV
1.1 mV
3.3 mV
+ Current
1
µ
A
48
µ
A
14
µ
A
160
µ
A
- Current
1
µ
A
37
µ
A
14
µ
A
151
µ
A
Fixed Overvoltage Protection:
(Measured at output terV and -V)
Minimum
56 V
56 V
56 V
56 V
Nominal
60 V
60 V
60 V
60 V
Maximum
64 V
64 V
64 V
64 V
Summary of Contents for 6625A
Page 10: ......
Page 24: ......
Page 30: ......
Page 44: ......
Page 62: ......
Page 68: ...Remote Operation 68 Figure 5 2 Sheet 1 of 2 Syntax Forms For Power Supply Commands ...
Page 69: ...Remote Operation 69 Figure 5 2 Sheet 2 of 2 Syntax Forms For Power Supply Commands ...
Page 90: ......
Page 94: ...Alignment Procedures 94 Figure A 3 Figure A 4 Figure A 5 ...
Page 116: ......
Page 120: ......