12
Installation Guide: Flame Ionization Detector (FID)
•
If this is a replacement FID, or is a second FID, then the
ignitor extension cable should be already installed. Go to
in the following procedure.
1
Route the 4-wire (two red and two white) ignitor extension
cable assembly along existing wiring and tubing located
behind the oven top:
•
Note that the flat, single connector is routed to the
electronics (right) side of the GC while the end with four
separate "push-on" connectors is routed to the flow
control (left) side.
•
Route the end with the flat, single connector to be located
below the previously-installed FID printed circuit board.
Route the end with the four separate "push-on" connectors
to be located at the rear of the previously-installed flow
control module.
2
Connect the flat, single connector on the ignitor extension
cable to its mating connector found on a "ribbon" cable
leading to the keyboard assembly.