14
Installation Guide: Flame Ionization Detector (FID)
•
If the previously-installed flow module is in the lower
detector position, select the white pair of wires (labeled
"BACK") and push each connector onto an electrical
terminal at the back of the flow module (no polarity).
In either case, there is no polarity concer; either connector
may be used on either terminal.
4
Use cable ties to secure the cable and previously-installed
FID tubing to existing wiring and tubing running behind the
oven top.
Installing the Insulation Cup
1
Install insulation into the cup.
2
Push the wire spring lever at the bottom of the cup to the
right to uncover the hole.