202
Agilent X-Series Signal Generators SCPI Command Reference
Analog Modulation Commands
Amplitude Modulation Subsystem–Option UNT ([:SOURce])
This command sets the couplings between start- to- stop values and stop- to- start values for AM sweep
times and rates. The sweep times coupled to on.
Use this command when
:AM[1]|2:INTernal:SWEep:SHAPe
is set to TRIangle.
ON|1
This choice uses the same sweep time for both start freq to stop freq and stop
freq to start freq sweeps of a triangle shape sweep. If sweep times coupled is ON,
sweep time and rate is only set by the Start - > Stop softkeys.
OFF|0
This choice disables sweep time couplings.
*RST
OFF
:AM[1]|2:INTernal:SWEep:TIME[1]|2
Supported
All Models with Options UNT and 303
[:SOURce]:AM[1]|2:INTernal:SWEep:TIME[1]|2 <val><unit>
[:SOURce]:AM[1]|2:INTernal:SWEep:TIME[1]|2?
This command sets the sweep rate for the amplitude- modulated, swept- sine waveform.
*RST
+1.00000000E
001
Range
1mS–65.535S
Key Entry
AM Sweep Time
:AM[1]|2:INTernal:SWEep:TRIGger
Supported
All Models with Options UNT and 303
[:SOURce]:AM[1]|2:INTernal:SWEep:TRIGger BUS|IMMediate|EXTernal|INTernal|KEY
[:SOURce]:AM[1]|2:INTernal:SWEep:TRIGger?
This command sets the trigger source for the amplitude modulated swept- sine waveform.
BUS
This choice enables GPIB triggering using the
*TRG
or
GET
command or LAN and
RS- 232 triggering using the
*TRG
command.
IMMediate
This choice enables immediate triggering of the sweep event.
EXTernal
This choice enables the triggering of a sweep event by an externally applied signal at the
TRIG 1, TRIG 2, or PULSE BNC connector.
INTernal
This choice enables the triggering of a sweep event by an internal Pulse Video or Pulse
Sync signal.
KEY
This choice enables triggering through front panel interaction by pressing the
Trigger
hardkey.
*RST
IMM
Key Entry
AM Sweep Times Coupled Off On
Key Entry
Bus
Free Run
Ext
Int
Trigger Key