47
Programming Reference
SCPI Instrument Command List
Command
:OUTP[1]
Long
:OUTPut[1][:NORMal][:STATe]
Form
Set & Query
Parameter
ON | OFF | 1 | 0
*RST value
OFF
Description
Use this command to switch the normal OUTPUT on or off.
Example
To switch on the output:
:OUTP ON
Command
:OUTP[1]:IMP
Long
:OUTPut[1]:IMPedance[:INTernal]
Form
Set & Query
Parameter
Numeric
Parameter Suffix
OHM
with engineering prefixes, e.g.:
MOHM
is Megaohms.
*RST value
50
Ω
Specified Limits
50
Ω
or 1 k
Ω
Description
Use this command to program the source impedance of the OUTPUT
connector. Note that only two settings are available. If you try to program
any other value, it will be rounded to one of the specified values.
Example
To program the source impedance:
Command
:OUTP[1]:IMP:EXT
Long
:OUTPut[1]:IMPedance:EXTernal
Form
Set & Query
Parameter
Numeric
Parameter Suffix
OHM
with engineering prefixes, e.g.:
MOHM
is Megaohms.
*RST value
50.0
Ω
:OUTP:IMP 50OHM
Set OUTPUT impedance to 50
Ω