Cleaning Information
Cleaning Instructions
102
Agilent
81950A Compact TLS module, First Edition
How to clean connector interfaces
Preferred Procedure
Use the following procedure on most occasions.
1
Clean the interface by pushing and pulling a new, dry pipe-cleaner into
the opening. Rotate the pipe-cleaner slowly as you do this.
2
Then clean the interface by rubbing a new, dry cotton-swab over the
surface using a small circular movement.
3
Blow away any remaining lint with compressed air.
Procedure for Stubborn Dirt
Use this procedure particularly when there is greasy dirt on the interface:
1
Moisten a new pipe-cleaner with isopropyl alcohol.
2
Clean the interface by pushing and pulling the pipe-cleaner into the
opening. Rotate the pipe-cleaner slowly as you do this.
3
Moisten a new cotton-swab with isopropyl alcohol.
4
Clean the interface by rubbing the cotton-swab over the surface using a
small circular movement.
5
Using a new, dry pipe-cleaner, and a new, dry cotton-swab remove the
alcohol, any dissolved sediment and dust.
6
Blow away any remaining lint with compressed air.
Be careful when using pipe-cleaners, as the core and the bristles of
the pipe-cleaner are hard and can damage the interface.
Do not use pipe-cleaners on optical head adapters, as the hard core
of normal pipe cleaners can damage the bottom of an adapter.
C A U T I O N
Summary of Contents for 81950A
Page 1: ...Agilent Technologies Agilent 81950A Compact Tunable Laser Source module User s Guide ...
Page 9: ...Agilent 81950A Compact TLS module First Edition 9 ...
Page 10: ...10 Agilent 81950A Compact TLS module First Edition ...
Page 24: ...Getting Started Signal Input and Output 24 Agilent 81950A Compact TLS module First Edition ...
Page 30: ...Accessories User s Guides 30 Agilent 81950A Compact TLS module First Edition ...
Page 50: ...Operating Instructions The User Interface 50 Agilent 81950A Compact TLS module First Edition ...