Operation
Operating Instructions
Agilent
81950A Compact TLS module, First Edition
33
Operation
Parameters
The instrument allows setting the following parameters:
•
Laser State
•
Optical Output Power
•
Operation Mode ("Auto Mode" or "Grid Mode")
•
Frequency, Wavelength
•
Reference Frequency
•
Grid Spacing
•
Channel Number
•
Frequency Offset (or "Fine Tune Frequency")
•
Modulation
•
State
•
SBS suppression FM frequency
•
Signaling or trace tone AM level
The AM modulation is only available with the Agilent 81950A module
using the BNC connector at the module's front panel. The N7711A /
N7714A instruments platform features no such connector.
Output Frequency
The output frequency is calculated using this equation:
Limitations
•
Reference frequency and grid spacing can only be changed when the
laser is off.
•
Operation mode can only be changed when the laser is off.
•
The maximum grid spacing value the underlying hardware can handle
is 3.2767 THz and depending on the specific instrument it may be less
than that.
f
fo
c
s
f
Δ
+
×
+
=