Definition of Terms
Specifications
Agilent
81950A Compact TLS module, First Edition
83
Measurement:
Using an Agilent Lightwave signal analyzer and bandwidth set to 3 MHz.
Relative wavelength (frequency) accuracy
When randomly changing the wavelength (frequency) of the tunable laser
source and measuring the differences between the displayed and the
actual wavelength (frequency), the relative wavelength (frequency)
accuracy is ± half the span between the maximum and the minimum value
of all differences.
Figure 9 Relative wavelength accuracy.
Conditions:
Uninterrupted tunable laser source output power, constant power setting,
constant temperature.
Return loss
Specifies the ratio of the optical power incident to the tunable laser source
output port at the wavelength set on the tunable laser source, to the power
reflected from the tunable laser source output port.
Conditions:
Tunable laser source output off.
Summary of Contents for 81950A
Page 1: ...Agilent Technologies Agilent 81950A Compact Tunable Laser Source module User s Guide ...
Page 9: ...Agilent 81950A Compact TLS module First Edition 9 ...
Page 10: ...10 Agilent 81950A Compact TLS module First Edition ...
Page 24: ...Getting Started Signal Input and Output 24 Agilent 81950A Compact TLS module First Edition ...
Page 30: ...Accessories User s Guides 30 Agilent 81950A Compact TLS module First Edition ...
Page 50: ...Operating Instructions The User Interface 50 Agilent 81950A Compact TLS module First Edition ...