3-3
Specifications and Regulatory Information
Specifications
Specifications
General Specifications
Total output power
a
+8.1 dBm maximum (6.5 mW)
+5.5 dBm minimum (3.5 mW)
+5.1 dBm maximum (3.2 mW) (Option 009)
+2.5 dBm minimum (1.8 mW) (Option 009)
Power stability
b
<
±
0.02 dB (15 minutes)
<
±
0.05 dB (6 hours)
Compatible fiber
9/125
µ
m, single mode
a. Measured with an InGaAs power sensor.
b. Ambient temperature change <
±
1
°
C measured with power meter having >30 dB return loss and after one hour warm-up time.
General Characteristics
Spectral density
1530 nm
1550 nm
1560 nm
> –13 dBm [1 nm] (> 50
µ
W/nm)
> –16 dBm [1 nm] (> 25
µ
W/nm) (Option 009)
> –13 dBm [1 nm] (> 50
µ
W/nm)
> –16 dBm [1 nm] (> 25
µ
W/nm) (Option 009)
> –13 dBm [1 nm] (> 50
µ
W/nm)
> –16 dBm [1 nm] (> 25
µ
W/nm) (Option 009)
Output return loss
> 30 dB (characteristic)
> 50 dB (characteristic)
a
Degree of polarization
< 5%, standard (characteristic)
> 95%, Option 009 (characteristic)
Modulation
Digital (TTL compatible)
, DC to 300 Hz (characteristic)
a. Measured at 1550 nm with FC/APC connector (Option 022).
Summary of Contents for 83437A
Page 1: ...Agilent 83438A Erbium ASE Source User s Guide ...
Page 5: ...v The Agilent 83438A At a Glance Rear view of instrument ...
Page 8: ......
Page 10: ......
Page 24: ...2 4 Making Measurements Performing Stimulus Response Measurements ...
Page 41: ...3 Specifications 3 3 Regulatory Information 3 6 Specifications and Regulatory Information ...
Page 48: ......
Page 54: ......
Page 61: ...5 7 Servicing General Information ...
Page 63: ...5 9 Servicing General Information ...
Page 79: ...5 25 Servicing Adjustment Procedure ...
Page 85: ...5 31 Servicing Replacing Instrument Assemblies Location of resistors R2 R8 and R9 ...
Page 92: ...5 38 Servicing Replaceable Parts ...
Page 94: ...5 40 Servicing Replaceable Parts ...
Page 96: ...5 42 Servicing Replaceable Parts ...
Page 98: ...5 44 Servicing Replaceable Parts ...
Page 100: ...5 46 Servicing Replaceable Parts ...
Page 106: ......