2-5
Making Measurements
Performing Stimulus-Response Measurements
To characterize a passive device
1
Turn on the optical spectrum analyzer and the Agilent 83438A, and allow them
to warm up for 1 hour.
2
Use a fiber optic cable to connect the Agilent 83438A’s output to the input of
the optical spectrum analyzer.
3
On the optical spectrum analyzer, press
INSTR PRESET
,
AUTO MEAS
,
and then
AUTO ALIGN
.
4
On the optical spectrum analyzer, use the
START
and
STOP
keys to set the proper
wavelength range.
5
Press
REF LEVEL
, and set the reference level to –10 dBm.
6
Press
SENS,
and set the sensitivity to –70 dBm. You can increase the sensitivity
beyond this value if needed.
7
If you are measuring a deep-notch filter, more sensitivity may be required.
Perform the following steps:
a
Press
Amptd
,
MORE
, and then
CHOP On
so that
On
is underlined.
b
Press
BW,Swp
,
SWPTIME AutoMan
, and set the sweep time to 50 seconds.
These steps cause the optical spectrum analyzer to compensate for the
effect of stray light inside the analyzer’s monochromator.
8
Press
State
,
instr modes
, and then
STM/RESP
.
9
Press
STORE THRU->B,
and then
NORM On Off
to normalize the response.
Summary of Contents for 83437A
Page 1: ...Agilent 83438A Erbium ASE Source User s Guide ...
Page 5: ...v The Agilent 83438A At a Glance Rear view of instrument ...
Page 8: ......
Page 10: ......
Page 24: ...2 4 Making Measurements Performing Stimulus Response Measurements ...
Page 41: ...3 Specifications 3 3 Regulatory Information 3 6 Specifications and Regulatory Information ...
Page 48: ......
Page 54: ......
Page 61: ...5 7 Servicing General Information ...
Page 63: ...5 9 Servicing General Information ...
Page 79: ...5 25 Servicing Adjustment Procedure ...
Page 85: ...5 31 Servicing Replacing Instrument Assemblies Location of resistors R2 R8 and R9 ...
Page 92: ...5 38 Servicing Replaceable Parts ...
Page 94: ...5 40 Servicing Replaceable Parts ...
Page 96: ...5 42 Servicing Replaceable Parts ...
Page 98: ...5 44 Servicing Replaceable Parts ...
Page 100: ...5 46 Servicing Replaceable Parts ...
Page 106: ......