4-12
8510XF Network Analyzer Systems
Performance Verification
Verification Procedures
High Band Verification
(> 50 GHz)
NOTE
Before verification of the high band frequencies it is necessary to: 1) select
new hardware for the system and 2) perform a calibration (for the frequency
band > 50 GHz).
1. From the Main Menu select
System Config
then
Hardware Config
.
2. Select the proper hardware setup for High Band Verification (> 50 GHz)
using
for the E7352A 110 GHz System or
3. Select
Done
and
Prior Menu
to return to Main Menu.
4. Select
Verify System
to go to the System Performance Verification Menu.
5. Select
System Cal
and follow the program prompts to calibrate the system
for frequencies >50 GHz. See
for the calibrations
required and suggested calibration storage locations (refer to
6. Select
Select Standard
to go to the Standard Selection Menu.
7. To verify the system use the Verification Standards provided with the
85059A 1.0 mm Precision Calibration and Verification Kit (refer to
). Using the menu select the verification standard
and enter the standard’s serial number. The verification standards may
be measured in any order.
NOTE
It is recommended to connect the Verification Standards to the cable on
port 2 prior to attempting to connect to port 1. Then follow the
recommended procedure for making a thru connection outlined under
to avoid damaging the cable.
8. Change default for
Cal Set for Broadband Measurement
from
1
to
3
.
NOTE
It is suggested that the High Band Calibration be stored in Cal Set 3 (Refer
to
9. Verify
Cal Set for 45 MHz Measurement (If Required)
is set to
No
.
10. Select
Done
and follow the program prompts.
11. Select
Measure Data
to measure the verification standards. Follow the
program prompts.
Summary of Contents for 8510XF
Page 11: ...8510XF Network Analyzer Systems xi Compliance with Standards ...
Page 12: ...xii 8510XF Network Analyzer Systems Compliance with Standards ...
Page 136: ...4 18 8510XF Network Analyzer Systems Performance Verification CW Frequency Accuracy Test ...
Page 160: ...5 24 8510XF Network Analyzer Systems System Maintenance System Block Diagrams ...
Page 162: ...5 26 8510XF Network Analyzer Systems System Maintenance System Block Diagrams ...
Page 164: ...5 28 8510XF Network Analyzer Systems System Maintenance System Block Diagrams ...
Page 168: ...5 32 8510XF Network Analyzer Systems System Maintenance System Block Diagrams ...
Page 170: ...5 34 8510XF Network Analyzer Systems System Maintenance System Block Diagrams ...
Page 172: ...5 36 8510XF Network Analyzer Systems System Maintenance System Block Diagrams ...
Page 176: ...5 40 8510XF Network Analyzer Systems System Maintenance Troubleshooting ...
Page 188: ...6 12 8510XF Network Analyzer Systems Replaceable Parts Millimeter Wave Controller ...
Page 191: ...R ...
Page 192: ...R ...
Page 193: ...R ...
Page 194: ...R ...
Page 195: ...R ...
Page 196: ...R ...
Page 197: ...R ...
Page 198: ...HP E7350 calmckit cdr Rev 1 3 10 98 Menu Maps Cal Modify Cal Kit 7 17 7 18 ...
Page 199: ...R ...
Page 200: ...R ...
Page 201: ......
Page 202: ...R ...
Page 203: ...R ...
Page 204: ...R ...
Page 205: ...R ...
Page 206: ...R ...
Page 207: ...R ...
Page 208: ...R ...
Page 209: ...R ...
Page 210: ...R ...