3-12 8510XF Network Analyzer Systems
Operation
Choosing Calibration Standards
Banded Standards
Some of the calibration standards in the 1.0 mm calibration kit are limited to
a particular frequency band. Calibration standards that are limited in this
way are shown in
below. During the calibration process, be sure to
use all the standards that are needed to cover the frequency range in which
you will be making measurements. Refer to the 85059A calibration kit
manual for more information.
Table 3-1
Banded Standards
∀
Non-Banded
Standards
Other standards in the 1.0 mm calibration kit, which are not limited to a
particular frequency range, are shown in
below. The right column
of the table refers to some devices that were also listed in
they are used differently. For example, Short No. 3 was limited to 50 GHz as
a “Shorts” standard, but not when it is used as a “Response” standard.
Table 3-2
Calibration Standards Used at All Frequencies
∀
Standard Type
.045 to 50 GHz
50 to 75 GHz
75 to 110 GHz
Combined Ranges
“Open/Short”
(M) OPEN
(F) OPEN
(M) SHORT3
(F) SHORT3
(M) SHORT3
(F) SHORT3
(combine the standards
shown in the columns to
the left, as appropriate)
“Shorts”
(M) SHORT3
(F) SHORT3
(M) SHORT1
(F) SHORT1
(M) SHORT1
(F) SHORT1
(combine the standards
shown in the columns to
the left, as appropriate)
“Load/Short”
50GHZ LOAD
(M) SHORT4
(F) SHORT4
(M) SHORT2
(F) SHORT2
(combine the standards
shown in the columns to
the left, as appropriate)
“Match” (TRL/TRM
Calibration)
50GHZ LOAD
(both ports)
N/A
N/A
(TRM calibration is not
used above 50 GHz)
“Isolation”
50 GHz LOAD
Load BB
1
1. The broadband load is a combination of a lossy delay line plus a 50 GHz load (refer to the 85059A calibration kit manual).
Load BB
1
Load BB
1
Standard Types
Device or Connection
“Response” (for S11 or S22)
Choose one of the following:
• (M) SHORT3 or (F) SHORT3
• (M) OPEN or (F) OPEN
“Response” (for S21 or S12)
• THRU (connect Port 1 to Port 2 via cable)
“S11REFLECT SHORT” (TRL menu)
Connect a short to Port 1:
(M) SHORT3 or (F) SHORT3
“S22REFLECT SHORT” (TRL menu)
Connect a short to Port 2:
(M) SHORT3 or (F) SHORT3
“THRU” (various menus)
Connect Port 1 to Port 2 via cable
Summary of Contents for 8510XF
Page 11: ...8510XF Network Analyzer Systems xi Compliance with Standards ...
Page 12: ...xii 8510XF Network Analyzer Systems Compliance with Standards ...
Page 136: ...4 18 8510XF Network Analyzer Systems Performance Verification CW Frequency Accuracy Test ...
Page 160: ...5 24 8510XF Network Analyzer Systems System Maintenance System Block Diagrams ...
Page 162: ...5 26 8510XF Network Analyzer Systems System Maintenance System Block Diagrams ...
Page 164: ...5 28 8510XF Network Analyzer Systems System Maintenance System Block Diagrams ...
Page 168: ...5 32 8510XF Network Analyzer Systems System Maintenance System Block Diagrams ...
Page 170: ...5 34 8510XF Network Analyzer Systems System Maintenance System Block Diagrams ...
Page 172: ...5 36 8510XF Network Analyzer Systems System Maintenance System Block Diagrams ...
Page 176: ...5 40 8510XF Network Analyzer Systems System Maintenance Troubleshooting ...
Page 188: ...6 12 8510XF Network Analyzer Systems Replaceable Parts Millimeter Wave Controller ...
Page 191: ...R ...
Page 192: ...R ...
Page 193: ...R ...
Page 194: ...R ...
Page 195: ...R ...
Page 196: ...R ...
Page 197: ...R ...
Page 198: ...HP E7350 calmckit cdr Rev 1 3 10 98 Menu Maps Cal Modify Cal Kit 7 17 7 18 ...
Page 199: ...R ...
Page 200: ...R ...
Page 201: ......
Page 202: ...R ...
Page 203: ...R ...
Page 204: ...R ...
Page 205: ...R ...
Page 206: ...R ...
Page 207: ...R ...
Page 208: ...R ...
Page 209: ...R ...
Page 210: ...R ...