4-4
8510XF Network Analyzer Systems
Performance Verification
Frequency ranges
For most of the menu items in the software, there are two or more selections
that pertain to the 8510XF. These menu selections are differentiated by:
A. The frequency range of the millimeter wave subsystem
❍
E7352A: 110 GHz subsystem
❍
E7342A: 85 GHz subsystem
❍
Option 005: 45MHz to 2 GHz
B. The portion of the subsystem’s frequency range that is being verified (the
ranges above and below 50 GHz are verified separately).
The menu selections are broken down in Tables 4-2 and 4-3, below.
Table 4-2
System Hardware Configuration (E7352A -- 110 GHz System)
Hardware
Selection (<50 GHz Range)
Selection (>50 GHz Range)
Network Analyzer
8510C -- Vector Network Analyzer
8510C -- Vector Network Analyzer
Test Set
E7352A -- 1.0 mm Subsystem (2-50 GHz)
E7352A005 -- 1.0 mm Subsystem (45MHz-50GHz)
E7352A110 -- 1.0 mm (50 GHz-110 GHz)
Source
8365xXF -- Synthesizer (45 MHz-50 GHz)
836xxXF -- Multiplied Synth. (above 50 GHz)
Calibration Kit
1
85059A -- 1.0 mm (45 MHz-50 GHz) [Use w/
E7342A/52A]
85059A110 -- 1.0 mm (50 GHz-110 GHz)
Calibration
Technique
BL -- Broadband Load Cal
OS -- Offset Short Cal
Test Port Cables
DIRECTXF -- for 8510XF verification (1.0 mm-1.0 mm)
DIRECTXF -- for 8510XF verification (1.0 mm-1.0 mm)
Verification Kit
85059AOV -- 1.0 mm (45 MHz-50 GHz)
85059AOV110 -- 1.0 mm (above 50 GHz)
1. The 85059A --1.0 mm (45MHz-50GHz) [Use w/ 8517A/B] selection is not used for the 8510XF Performance Verification.
Table 4-3
System Hardware Configuration (E7342A -- 85GHz System)
Hardware
Selection (<50 GHz Range)
Selection (>50 GHz Range)
Network Analyzer
8510C -- Vector Network Analyzer
8510C -- Vector Network Analyzer
Test Set
E7342A -- 1.0 mm Subsystem (2-50 GHz)
E7342A005 -- 1.0 mm Subsystem (45 MHz-50GHz)
E7342A85 -- 1.0 mm (50 GHz-85 GHz)
Source
8365xXF -- Synthesizer (45 MHz-50 GHz)
836xxXF --
Synth. (above 50 GHz)
Calibration Kit
1
85059A -- 1.0 mm (45 MHz-50 GHz) [Use w/
E7342A/52A]
85059A85 -- 1.0 mm (50 GHz-85 GHz)
Calibration
Technique
BL -- Broadband Load Cal
OS -- Offset Short Cal
Test Port Cables
DIRECTXF -- for 8510XF verification (1.0 mm-1.0 mm)
DIRECTXF -- for 8510XF verification (1.0 mm-1.0 mm)
Verification Kit
85059AOV -- 1.0 mm (45 MHz-50 GHz)
85059AOV110 -- 1.0 mm (above 50 GHz)
1. The 85059A --1.0 mm (45MHz-50GHz) [Use w/ 8517A/B] selection is not used for the 8510XF Performance Verification.
Summary of Contents for 8510XF
Page 11: ...8510XF Network Analyzer Systems xi Compliance with Standards ...
Page 12: ...xii 8510XF Network Analyzer Systems Compliance with Standards ...
Page 136: ...4 18 8510XF Network Analyzer Systems Performance Verification CW Frequency Accuracy Test ...
Page 160: ...5 24 8510XF Network Analyzer Systems System Maintenance System Block Diagrams ...
Page 162: ...5 26 8510XF Network Analyzer Systems System Maintenance System Block Diagrams ...
Page 164: ...5 28 8510XF Network Analyzer Systems System Maintenance System Block Diagrams ...
Page 168: ...5 32 8510XF Network Analyzer Systems System Maintenance System Block Diagrams ...
Page 170: ...5 34 8510XF Network Analyzer Systems System Maintenance System Block Diagrams ...
Page 172: ...5 36 8510XF Network Analyzer Systems System Maintenance System Block Diagrams ...
Page 176: ...5 40 8510XF Network Analyzer Systems System Maintenance Troubleshooting ...
Page 188: ...6 12 8510XF Network Analyzer Systems Replaceable Parts Millimeter Wave Controller ...
Page 191: ...R ...
Page 192: ...R ...
Page 193: ...R ...
Page 194: ...R ...
Page 195: ...R ...
Page 196: ...R ...
Page 197: ...R ...
Page 198: ...HP E7350 calmckit cdr Rev 1 3 10 98 Menu Maps Cal Modify Cal Kit 7 17 7 18 ...
Page 199: ...R ...
Page 200: ...R ...
Page 201: ......
Page 202: ...R ...
Page 203: ...R ...
Page 204: ...R ...
Page 205: ...R ...
Page 206: ...R ...
Page 207: ...R ...
Page 208: ...R ...
Page 209: ...R ...
Page 210: ...R ...