8510XF Network Analyzer Systems 4-9
Performance Verification
Verification Procedures
Verification Procedures
Low Band Verification
(< 50 GHz)
The following steps assume the hardware configuration has been previously
setup for Low Band Verification (< 50 GHz) in
Software Configuration.
1. From the Main Menu select
Verify System
to go to the System Performance
Verification Menu.
2. Select
Serial Numbers
to enter serial numbers of system and components.
3. Select
System Cal
and follow the program prompts to calibrate the system
for frequencies <50 GHz. See
for the calibrations
required and suggested calibration storage locations (also refer to the
“Two-Band Calibration Procedure for 8510XF Performance
Verification” foldout at the end of this chapter).
a.
Install the 1.0 mm female to female cable (8.8 cm, Agilent part
number 11500-60001) onto the right test head (port 2) prior to
calibration. This cable is part of the 85059A 1.0 mm Precision
Calibration and Verification Kit and is considered to be a test port
cable.
b.
A full two-port calibration is required. A table of calibration
standards required for each frequency range can be found in the
85059A 1.0 mm precision calibration and verification kit manual, or
on the foldout at the end of this chapter.
CAUTION
Do not remove this test port cable once the Calibration/Verification process
has begun. If the test port cable becomes loose or is removed during the
calibration/verification process, the calibration is invalid.
CAUTION
Do not pull on the connectors on the ends of the test port cable —this will
damage the cable. Avoid damaging the cable once it is connected to port 2
by carefully making the thru connection between the two test heads. This is
best done by placing the two test heads on a smooth surface and gently
sliding the right test head (port 2) towards the left test head (port 1).
Summary of Contents for 8510XF
Page 11: ...8510XF Network Analyzer Systems xi Compliance with Standards ...
Page 12: ...xii 8510XF Network Analyzer Systems Compliance with Standards ...
Page 136: ...4 18 8510XF Network Analyzer Systems Performance Verification CW Frequency Accuracy Test ...
Page 160: ...5 24 8510XF Network Analyzer Systems System Maintenance System Block Diagrams ...
Page 162: ...5 26 8510XF Network Analyzer Systems System Maintenance System Block Diagrams ...
Page 164: ...5 28 8510XF Network Analyzer Systems System Maintenance System Block Diagrams ...
Page 168: ...5 32 8510XF Network Analyzer Systems System Maintenance System Block Diagrams ...
Page 170: ...5 34 8510XF Network Analyzer Systems System Maintenance System Block Diagrams ...
Page 172: ...5 36 8510XF Network Analyzer Systems System Maintenance System Block Diagrams ...
Page 176: ...5 40 8510XF Network Analyzer Systems System Maintenance Troubleshooting ...
Page 188: ...6 12 8510XF Network Analyzer Systems Replaceable Parts Millimeter Wave Controller ...
Page 191: ...R ...
Page 192: ...R ...
Page 193: ...R ...
Page 194: ...R ...
Page 195: ...R ...
Page 196: ...R ...
Page 197: ...R ...
Page 198: ...HP E7350 calmckit cdr Rev 1 3 10 98 Menu Maps Cal Modify Cal Kit 7 17 7 18 ...
Page 199: ...R ...
Page 200: ...R ...
Page 201: ......
Page 202: ...R ...
Page 203: ...R ...
Page 204: ...R ...
Page 205: ...R ...
Page 206: ...R ...
Page 207: ...R ...
Page 208: ...R ...
Page 209: ...R ...
Page 210: ...R ...