3-6 8510XF Network Analyzer Systems
Operation
Measurement Calibration
Measurement Calibration
Why Calibration Is
Essential
During measurement calibration, a standard device with known
characteristics is measured, and the results of the measurement are used to
create a set of error coefficients. During subsequent measurements, the
network analyzer uses the error coefficients to correct for its own
measurement errors.
The accuracy of the network analyzer is very much dependent on this
error-correction technique; therefore, the analyzer must be calibrated before
it is used to make measurements.
When to Repeat the
Calibration
The 8510XF is designed for exceptional stability; after it has been
calibrated, the specified measurement accuracy of the system is guaranteed,
provided that the conditions of measurement remain the same. However,
certain changes to the measurement conditions may invalidate the
calibration, as described below.
•
A calibration is only valid for the frequency range over which it was
performed; if the frequency range has changed, the calibration must be
repeated.
•
A calibration is only valid for the number of points over which it was
performed; if the number of points has been increased, the calibration
must be repeated.
•
If test port cables have been kinked, replaced, or disconnected and
reconnected, the calibration may be invalid, and should be checked.
•
If connectors have been degraded by wear, the calibration may be
invalid, and should be checked.
•
If the ambient temperature changes by
±
1
°
C or more, the calibration
may be invalid, and should be checked.
Checking the
Calibration
A common practice for making a rough check of a calibration is to
remeasure one of the devices used in the calibration procedure (or
verification procedure). For example, if you reconnect a short that was used
during calibration, it should appear as a near-perfect short, across the
frequency range of calibration.
Summary of Contents for 8510XF
Page 11: ...8510XF Network Analyzer Systems xi Compliance with Standards ...
Page 12: ...xii 8510XF Network Analyzer Systems Compliance with Standards ...
Page 136: ...4 18 8510XF Network Analyzer Systems Performance Verification CW Frequency Accuracy Test ...
Page 160: ...5 24 8510XF Network Analyzer Systems System Maintenance System Block Diagrams ...
Page 162: ...5 26 8510XF Network Analyzer Systems System Maintenance System Block Diagrams ...
Page 164: ...5 28 8510XF Network Analyzer Systems System Maintenance System Block Diagrams ...
Page 168: ...5 32 8510XF Network Analyzer Systems System Maintenance System Block Diagrams ...
Page 170: ...5 34 8510XF Network Analyzer Systems System Maintenance System Block Diagrams ...
Page 172: ...5 36 8510XF Network Analyzer Systems System Maintenance System Block Diagrams ...
Page 176: ...5 40 8510XF Network Analyzer Systems System Maintenance Troubleshooting ...
Page 188: ...6 12 8510XF Network Analyzer Systems Replaceable Parts Millimeter Wave Controller ...
Page 191: ...R ...
Page 192: ...R ...
Page 193: ...R ...
Page 194: ...R ...
Page 195: ...R ...
Page 196: ...R ...
Page 197: ...R ...
Page 198: ...HP E7350 calmckit cdr Rev 1 3 10 98 Menu Maps Cal Modify Cal Kit 7 17 7 18 ...
Page 199: ...R ...
Page 200: ...R ...
Page 201: ......
Page 202: ...R ...
Page 203: ...R ...
Page 204: ...R ...
Page 205: ...R ...
Page 206: ...R ...
Page 207: ...R ...
Page 208: ...R ...
Page 209: ...R ...
Page 210: ...R ...