Enhancing Measurement Accuracy
A
Installation and User’s Guide
117
Periodic System Component Calibration
A complete calibration verifies that the system components meet their
individual performance specifications.
The calibration interval depends on the level of system use. Agilent
Technologies recommends an initial cycle of 6 to 12 months. Thereafter,
adjust the cycle based on the recalibration results.
A complete system calibration consists of the following tests:
•
Agilent E8364B PNA Series vector network analyzer performance tests
*
•
Agilent 4156C precision semiconductor parameter analyzer performance
verification
†
•
Agilent E5260A/70B precision parametric measurement solution
performance verification
•
Agilent 4284A precision LCR meter performance tests
•
Agilent 35670A dynamic signal analyzer performance verification
•
Agilent E5250A low leakage switch mainframe performance tests
See the individual instrument documentation for instructions on
performing the required instrument verification tests.
* Refer to the Agilent E8364B PNA Series vector network analyzer service guide for required performance tests.
The service guide (part number E8364-90026) is available at www.agilent.com in PDF format.
† Refer to the chapter titled “Performance Verification” in the Agilent 4156C precision semiconductor
parameter analyzer service guide, included with the Agilent 4156C documentation.
Summary of Contents for 85225F
Page 1: ...Agilent Technologies Agilent 85225F PerformanceModelingSystem Installation and User s Guide ...
Page 90: ...90 Installation and User s Guide 2 Installing the System ...
Page 102: ...102 Installation and User s Guide 3 Verifying System Functionality ...
Page 110: ...110 Installation and User s Guide 4 Servicing the System ...
Page 118: ...118 Installation and User s Guide A Enhancing Measurement Accuracy ...
Page 128: ...128 Installation and User s Guide C RF Subsystem Functional Verification Test ...