Chapter 2
155
Making Measurements
Example 12: Making Time-Domain Measurements with Sweep Delay
Unless the delay sweep function is used, the sweep starts immediately
after a valid trigger and lasts as long as the sweep time setting
indicates, and the only adjustment that can be made is to increase or
decrease the length of the sweep. A spectrum analyzer can show a
zero-span display, but if the spectrum analyzer is not using a sweep
delay, the amount of detail that can be seen for events that do not occur
immediately after the trigger event is limited. See Figure 2-81 for an
example of a zero-span display of an RF pulse that occurs 48 ms after
the external sweep trigger signal.
Summary of Contents for 8560E
Page 21: ...21 1 Quick Start Guide ...
Page 48: ...48 Chapter1 Quick Start Guide Manuals Available Separately ...
Page 49: ...49 2 Making Measurements ...
Page 164: ...164 Chapter2 Making Measurements Example 13 Making Pulsed RF Measurements ...
Page 165: ...165 3 Softkey Menus ...
Page 182: ...182 Chapter3 Softkey Menus Menu Trees ...
Page 183: ...183 4 Key Function Descriptions ...
Page 228: ...228 Chapter4 Key Function Descriptions Key Descriptions Figure 4 3 CRT Alignment Pattern ...
Page 289: ...289 5 Programming ...
Page 323: ...Chapter 5 323 Programming Math Functions Figure 5 10 Display Units ...
Page 339: ...339 6 Programming Command Cross Reference ...
Page 369: ...369 7 Language Reference ...
Page 654: ...654 Chapter7 Language Reference VTL Video Trigger Level ...
Page 655: ...655 8 Options and Accessories ...
Page 665: ...665 9 If You Have a Problem ...
Page 674: ...674 Chapter9 If You Have a Problem Power Requirements Figure 9 2 AC Power Cables Available ...