216
Chapter 4
Key Function Descriptions
Key Descriptions
A
W
appears on the left edge of the display to indicate
the function is active. If you have not previously edited
or recalled correction data, then the ampcor function is
not activated and the message
No Correction Loaded
appears in the active annunciator area of the display.
Each correction point consists of a frequency at which
the correction should be applied and an amplitude, in
dB, of the correction. For frequencies between
correction points, the ampcor function linearly
interpolates the correction data you have entered.
Frequencies outside the range over which you have
specified correction are corrected by the amplitude
values at the nearest (in frequency) end of the list of
correction data. The correction is automatically
recalculated for any span or vertical scale, including
linear.
If any of the trace data is above or below the graticule,
the ampcor function may not properly correct it. The
spectrum analyzer amplitude accuracy is not specified
above or below the graticule. Anytime ampcor applies a
correction such that data outside the graticule is moved
to within the graticule, an error (error number 921 or
922) is generated. In order to avoid these errors, make
sure that the trace data which is being corrected is
within the graticule before the correction is applied.
Front-panel key access:
CAL
AMPLITUDE
Accesses the amplitude functions, which are as follows:
REF LVL, NORM REF LVL, ATTEN AUTO MAN, LOG
dB/DIV, LINEAR, RANGE LVL, REF LVL OFFSET,
MAX MXR LEVEL, AMPTD UNITS, COUPLING AC
DC, NORM REF POSN, PRESEL AUTO PK, and
PRESEL MAN ADJ.
AMPLITUDE
also activates the
reference-level function.
Front-panel key access:
AMPLITUDE
AMPTD CORRECT
Displays a menu of functions that set conversion
losses and flatness data.
Front-panel key access:
AUX CTRL
AMPTD UNITS
Accesses the softkeys that allow you to change the
amplitude units of the spectrum analyzer. AMPTD
UNITS accesses following softkeys: UNITS AUTO
MAN, dBm, dB
µ
V, dBmV, VOLTS, and WATTS.
Front-panel key access:
AMPLITUDE
Summary of Contents for 8560E
Page 21: ...21 1 Quick Start Guide ...
Page 48: ...48 Chapter1 Quick Start Guide Manuals Available Separately ...
Page 49: ...49 2 Making Measurements ...
Page 164: ...164 Chapter2 Making Measurements Example 13 Making Pulsed RF Measurements ...
Page 165: ...165 3 Softkey Menus ...
Page 182: ...182 Chapter3 Softkey Menus Menu Trees ...
Page 183: ...183 4 Key Function Descriptions ...
Page 228: ...228 Chapter4 Key Function Descriptions Key Descriptions Figure 4 3 CRT Alignment Pattern ...
Page 289: ...289 5 Programming ...
Page 323: ...Chapter 5 323 Programming Math Functions Figure 5 10 Display Units ...
Page 339: ...339 6 Programming Command Cross Reference ...
Page 369: ...369 7 Language Reference ...
Page 654: ...654 Chapter7 Language Reference VTL Video Trigger Level ...
Page 655: ...655 8 Options and Accessories ...
Page 665: ...665 9 If You Have a Problem ...
Page 674: ...674 Chapter9 If You Have a Problem Power Requirements Figure 9 2 AC Power Cables Available ...