302
Chapter 5
Programming
Program Timing
Figure 5-7
Update trace with TS before executing marker commands.
As the example shows, TS is executed after analyzer settings are
changed, but before trace information is processed.
There are two commands that change the measurement range
indirectly: MKCF (marker to center frequency) and MKRL (marker to
reference level). They set the center frequency and reference level equal
to the marker frequency and amplitude, respectively. If valid trace
information is needed, a TS command must be executed after these
commands are executed.
The TS command invokes at least one entire sweep. It invokes more
than one sweep when certain commands are active, such as SP
(frequency span) when MKTRACK (marker tracking) is turned on.
NOTE
To ensure that all processes have been completed (for example, MKFC),
take a sweep and then query the data.
Summary of Contents for 8560E
Page 21: ...21 1 Quick Start Guide ...
Page 48: ...48 Chapter1 Quick Start Guide Manuals Available Separately ...
Page 49: ...49 2 Making Measurements ...
Page 164: ...164 Chapter2 Making Measurements Example 13 Making Pulsed RF Measurements ...
Page 165: ...165 3 Softkey Menus ...
Page 182: ...182 Chapter3 Softkey Menus Menu Trees ...
Page 183: ...183 4 Key Function Descriptions ...
Page 228: ...228 Chapter4 Key Function Descriptions Key Descriptions Figure 4 3 CRT Alignment Pattern ...
Page 289: ...289 5 Programming ...
Page 323: ...Chapter 5 323 Programming Math Functions Figure 5 10 Display Units ...
Page 339: ...339 6 Programming Command Cross Reference ...
Page 369: ...369 7 Language Reference ...
Page 654: ...654 Chapter7 Language Reference VTL Video Trigger Level ...
Page 655: ...655 8 Options and Accessories ...
Page 665: ...665 9 If You Have a Problem ...
Page 674: ...674 Chapter9 If You Have a Problem Power Requirements Figure 9 2 AC Power Cables Available ...