Chapter 7
421
Language Reference
AMPCORSAVE Amplitude Correction Save
AMPCORSAVE Amplitude Correction Save
Syntax
Figure 7-56
AMPCORSAVE Syntax
Description
The AMPCORSAVE function saves the current correction points in one
of ten possible registers. The correction points can be recalled with the
AMPCORRCL command.
Parameters
number
integer from 0 to 5
Example
10
REAL Regnumber
20
Regnumber=2
30
OUTPUT 718;"AMPCORSAVE";Regnumber;";"
40
END
Summary of Contents for 8560E
Page 21: ...21 1 Quick Start Guide ...
Page 48: ...48 Chapter1 Quick Start Guide Manuals Available Separately ...
Page 49: ...49 2 Making Measurements ...
Page 164: ...164 Chapter2 Making Measurements Example 13 Making Pulsed RF Measurements ...
Page 165: ...165 3 Softkey Menus ...
Page 182: ...182 Chapter3 Softkey Menus Menu Trees ...
Page 183: ...183 4 Key Function Descriptions ...
Page 228: ...228 Chapter4 Key Function Descriptions Key Descriptions Figure 4 3 CRT Alignment Pattern ...
Page 289: ...289 5 Programming ...
Page 323: ...Chapter 5 323 Programming Math Functions Figure 5 10 Display Units ...
Page 339: ...339 6 Programming Command Cross Reference ...
Page 369: ...369 7 Language Reference ...
Page 654: ...654 Chapter7 Language Reference VTL Video Trigger Level ...
Page 655: ...655 8 Options and Accessories ...
Page 665: ...665 9 If You Have a Problem ...
Page 674: ...674 Chapter9 If You Have a Problem Power Requirements Figure 9 2 AC Power Cables Available ...