AT Input Attenuation
AT Input Attenuation
Syntax
Figure 7-60
AT Syntax
Description
The AT command sets the amount of attenuation between the input
and the first mixer. The attenuation can be set to 0 dB only by numeric
data entry, and not by using the knob or step keys.
Parameters
number
integer from 0 to 70, in decade increments. Numbers
are rounded up to the nearest decade. (integer from 0 to
60 for the Agilent 8564E/EC and Agilent 8565E/EC)
AUTO
sets the attenuation to coupled mode.
MAN
sets the attenuation to manual mode.
UP or DN
changes the attenuation by 10 dB (but cannot set
attenuation to 0 dB).
Preset State
Coupled mode, 10 dB
Summary of Contents for 8560E
Page 21: ...21 1 Quick Start Guide ...
Page 48: ...48 Chapter1 Quick Start Guide Manuals Available Separately ...
Page 49: ...49 2 Making Measurements ...
Page 164: ...164 Chapter2 Making Measurements Example 13 Making Pulsed RF Measurements ...
Page 165: ...165 3 Softkey Menus ...
Page 182: ...182 Chapter3 Softkey Menus Menu Trees ...
Page 183: ...183 4 Key Function Descriptions ...
Page 228: ...228 Chapter4 Key Function Descriptions Key Descriptions Figure 4 3 CRT Alignment Pattern ...
Page 289: ...289 5 Programming ...
Page 323: ...Chapter 5 323 Programming Math Functions Figure 5 10 Display Units ...
Page 339: ...339 6 Programming Command Cross Reference ...
Page 369: ...369 7 Language Reference ...
Page 654: ...654 Chapter7 Language Reference VTL Video Trigger Level ...
Page 655: ...655 8 Options and Accessories ...
Page 665: ...665 9 If You Have a Problem ...
Page 674: ...674 Chapter9 If You Have a Problem Power Requirements Figure 9 2 AC Power Cables Available ...