648
Chapter 14
Analyzer Messages
Interpreting Analyzer Messages
4. When the comb-generator output amplifier is enabled, the
amplitude of the 100 MHz signal and the 100 MHz harmonics
increase. The 100 MHz signal amplitude increases to
approxi12 dBm.
❏
If the signal amplitude does not increase, check that the
comb-generator control line from the A7 analog interface
assembly is switching correctly.
Refer to Foldout 11-2, 11-3, or 11-4, RF Section Overall Block
Diagram for your analyzer, for more information about the
comb-generator control voltage values.
❏
If the A7 control line is correct, suspect the A3A11
step-recovery diode or the A3A1 comb generator.
Refer to Foldout 11-2, 11-3, or 11-4, RF Section Overall Block Diagram
for your analyzer, for more information.
OVEN COLD (M)
This message is displayed for instruments equipped with Option 004,
precision frequency reference.
The message is displayed for 5 minutes after the instrument is turned
on. The message alerts the user that the oven in the A22 precision
frequency reference has not been on long enough to warm the reference
to its operating temperature.
Related Assemblies
A22 precision frequency reference.
Troubleshooting Hints
This is a timed message that comes on whenever the instrument is
turned on; the temperature of the A22 precision frequency reference
oven-controlled crystal oscillator (OCXO) is not measured.
The instrument firmware displays the message only when it senses
that the A22 assembly is connected to its power supply through W15,
the OCXO power cable. If the message does not appear, check the W15
cable.
PARAMETER ERROR: _ _ _ (U)
The specified parameter is not recognized by the analyzer. See the
appropriate programming command description to determine the
correct parameters.
PASSCODE NEEDED (U)
Indicates that the function cannot be accessed without the pass code.
Summary of Contents for 8590L
Page 29: ...30 Chapter1 Introduction If You Replace or Repair an Assembly ...
Page 175: ...176 Chapter2 Making Adjustments 28 Correcting for Absolute Amplitude Accuracy for Option 051 ...
Page 279: ...280 Chapter5 Troubleshooting the IF Section Troubleshooting the A15 Motherboard Assembly ...
Page 347: ...348 Chapter7 Replacing Major Assemblies Rear Frame Assembly ...
Page 355: ...356 Chapter8 If You Have A Problem Returning Your Analyzer for Service ...
Page 379: ...380 Chapter9 Assembly Descriptions and Block Diagrams IF Section ...
Page 380: ......
Page 381: ...382 Chapter9 Assembly Descriptions and Block Diagrams IF Section ...
Page 382: ......
Page 383: ...384 Chapter9 Assembly Descriptions and Block Diagrams IF Section ...
Page 384: ......
Page 385: ...386 Chapter9 Assembly Descriptions and Block Diagrams IF Section ...
Page 386: ......
Page 387: ...388 Chapter9 Assembly Descriptions and Block Diagrams IF Section ...
Page 388: ......
Page 389: ...390 Chapter9 Assembly Descriptions and Block Diagrams IF Section ...
Page 390: ......
Page 391: ...392 Chapter9 Assembly Descriptions and Block Diagrams IF Section ...
Page 392: ...393 10 Analyzer Options and Block Diagrams ...
Page 443: ...444 Chapter10 Analyzer Options and Block Diagrams Option 190 DVB C Service Menus ...
Page 444: ......
Page 445: ...446 Chapter10 Analyzer Options and Block Diagrams Option 190 DVB C Service Menus ...
Page 446: ......
Page 447: ...448 Chapter10 Analyzer Options and Block Diagrams Option 190 DVB C Service Menus ...
Page 448: ......
Page 449: ...450 Chapter10 Analyzer Options and Block Diagrams Option 190 DVB C Service Menus ...
Page 451: ...452 Chapter10 Analyzer Options and Block Diagrams Option 190 DVB C Service Menus ...
Page 483: ...484 Chapter11 Major Assembly and Cable Locations Major Assembly and Cable Locations ...
Page 537: ...538 Chapter12 Replaceable Parts Component Level Information Packets ...
Page 538: ......
Page 539: ...540 Chapter12 Replaceable Parts Component Level Information Packets ...
Page 540: ......
Page 541: ...542 Chapter12 Replaceable Parts Component Level Information Packets ...
Page 542: ...FOLDOUT 12 3 8590L AND 8591E RF ASSEMBLY ...
Page 543: ...544 Chapter12 Replaceable Parts Component Level Information Packets ...
Page 544: ......
Page 545: ...546 Chapter12 Replaceable Parts Component Level Information Packets ...
Page 546: ......
Page 547: ...548 Chapter12 Replaceable Parts Component Level Information Packets ...
Page 548: ......
Page 549: ...550 Chapter12 Replaceable Parts Component Level Information Packets ...
Page 550: ......
Page 551: ...552 Chapter12 Replaceable Parts Component Level Information Packets ...
Page 665: ...666 Chapter15 Service Equipment and Tools Service Equipment and Tools ...
Page 666: ...667 16 Safety and Regulatory Information ...
Page 673: ...674 Chapter16 Safety and Regulatory Information Protection from Electrostatic Discharge ...