Agilent 86038B Photonic Dispersion and Loss Analyzer,
Second Edition
213
Server to Client Events
The server to client events allow the 86038B PDLA to update the
client with current data or instrument settings. In this context,
the PDLA is the server and your program is the client. These
events are asynchronous. That is, you cannot determine when
they are going to occur. You need to set up event handlers for
the events you want to catch.
Connected
(ByVal connected As Boolean)
Indicates that a remote session has connected to/disconnected
from the PDLA. Once you have requested a connection using the
Connect method on the Connectivity object, you must wait for
this event to return a true before sending commands to the
PDLA. If you receive this event after your connection has been
established, and the value is false, you need to re-establish your
connection.
Parameters
connected - True = Connected to the PDLA
Related Commands
• Connectivity.Connect method
• Connectivity.Disconnect method
• Connectivity.TerminateSession method
EstimatedTestTime
(ByVal sTestName As String, ByVal tsEstimate As Double)
Sends estimated time of a particular test. This event is sent once
at the beginning of a test. An action such as a Normalization
may be comprised of several tests. In this case, the
EstimatedTestTime event will be sent at the beginning of each
test.
Parameters
sTestName - The name of the test.
tsEstimate - The estimated time of the test in seconds.
Summary of Contents for 86038B
Page 1: ...Agilent 86038B Photonic Dispersion and Loss Analyzer User s Guide ...
Page 4: ...4 ...
Page 20: ...20 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition ...
Page 34: ...34 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition Figure 2 b Rear Panel ...
Page 78: ...78 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition ...
Page 92: ...92 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition ...
Page 202: ...202 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition End Sub ...
Page 348: ...348 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition ...
Page 349: ......