238
Agilent 86038B Photonic Dispersion and Loss Analyzer, Second Edition
ByVal measurementType As
InstrumentObjects.ODACommon.eMeasurementType - The type
of measurement to display: CD, Gain/Loss, GD, PDL, or DGD.
VB.NET Syntax
pdlaClient.Display.DefineTrace(0, _
InstrumentObjects.ODACommon.eTraceLetter.TraceA, _
InstrumentObjects.ODACommon.eODAPort.One, _
InstrumentObjects.ODACommon.eMeasurementType.GD)
VB 6.0 Syntax
Call pdlaClient.Display.DefineTrace(0, _
eTraceLetter_TraceA, _
eODAPort_One, _
eMeasurementType_GD)
GDOffsetMode Property
The PDLA normally sets the group delay at the center trace
point to zero. Writing a value to this property can change the
point that is zeroed to the minimum or maximum of the trace.
Reading the value of the property returns the current setting.
Property Value
InstrumentObjects.ODASensitivity.eGDOffsetMode
VB.NET Syntax
pdlaClient.Display.GDOffsetMode = _
InstrumentObjects.ODASensitivity.eGDOffsetMode.CenterPoint
VB 6.0 Syntax
pdlaClient.Display.GDOffsetMode =
eGDOffsetMode_CenterPoint
ToggleMode Method
Toggles the PDLA user interface graph display between one,
two, and four graphs. This method corresponds to the Change
View front panel key.
Summary of Contents for 86038B
Page 1: ...Agilent 86038B Photonic Dispersion and Loss Analyzer User s Guide ...
Page 4: ...4 ...
Page 20: ...20 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition ...
Page 34: ...34 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition Figure 2 b Rear Panel ...
Page 78: ...78 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition ...
Page 92: ...92 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition ...
Page 202: ...202 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition End Sub ...
Page 348: ...348 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition ...
Page 349: ......