Agilent 86038B Photonic Dispersion and Loss Analyzer, Second Edition
Figure 8
Sample CD trace for the MXF fiber
c By pressing Begin Length, the length can be measured.
This should be about 2 m. (This value is not calibrated for
the device under test). Then the CD measurement can be
displayed as CD Coefficient in ps/nm-km. as in Fig. 9. The
value shown should be similar to the values for standard
single-mode fiber. In particular it should show a value
near 17 ps/nm-km in the range from 1550-1560 nm.