Length Normalization
If a Length measurement is specified with the Measure mode,
the Normalization path’s length is measured so that it can be
subtracted out.
DUT Length = Total Length – Normalized Path Length.
to setup a length measurement.
Wavelength Accuracy Correction
Using the optional 86122A Multi-Wavelength Meter, wavelength
correction is performed for stepped normalizations, when a
stepped measurement is chosen.
Wavelength correction is then always performed if an 86122A is
connected. Five wavelength points are measured and the
average offset is applied. For more information, refer to