2-53
Function Reference
Scale/Div
Scale/Div
Specifies the dB per division of the vertical (that is, amplitude) scale. If Auto
is specified, the measurement function sets the vertical scale based on the
dynamic range of the signal. If a particular vertical scale is desired, clear the
Auto check box and put the desired dB/div in the dB dialog box. The ampli-
tude scale can be set from 0.01 to 20 dB per division. The Preset value is 10 dB
per division. The default selection is Auto.
Key Path
Amplitude > Scale/Div
Related Functions
Display Mode Log Linear, Reference Level
Remote
Commands
DISPlay:WINDow:TRACe:Y:SCALe:PDIVision
Search Limit On/Off
When the search limit function is on, all the marker peak/pit searches will
apply only to the range specified by the two wavelength line markers.
Key Path
Markers > More Marker Functions > Line Marker Menu > Search Limit On/
Off
Remote
Commands
CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:SRANge:STATe
Search Mode Peak/Pit
Allows you to select between peak and pit search modes. The selections on
the Marker Search Menu will change depending on which of the two modes is
selected.
Key Path
Markers > More Marker Functions > Marker Search Menu > Search Mode
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com