3-34
Using the Analyzer
Measurement Techniques
1
From the front panel press
Amplitude
>
Reference Level.
Adjust the
reference level so the highest value in the trace just intersects the top graticule
(grid line) of the display.
2
Adjust
Scale/Div
so the desired response is visible and covers more than half
the display.
3
Press
Sweep
>
Repeat Sweep On
. Make sure that auto-sensitivity is on:
Amplitude
>
Sensitivity
[Auto].
4
Place a marker on the lowest amplitude feature of the trace you would like to
characterize using Markers. Note the sweep-to-sweep variation.
5
If the sweep-to-sweep variation is too high for your needs, increase the
sensitivity until the sweep-to-sweep variation is acceptable. The sweep will
slow down and multiple sweeps per trace update will start to occur as
sensitivity is increased.