9-11
Specifications and Regulatory Information
Amplitude Specifications
Table 9-2. Amplitude Specifications
a
Precision Sweep
Mode
Normal Sweep
Mode
Fast Sweep
Mode
Minimum DUT Bandwidth
2 pm
4 pm
14 pm
Total Accuracy
b
,
c
<10 dB insertion loss
d
Full Range
±
0.05 dB
±
0.09 dB
±
0.07 dB
±
0.10 dB
±
0.09 dB
±
0.11 dB
Amplitude Stability
d
Sweep to sweep
Continuous over
±
4
°
C
temperature swing
±
0.015 dB
±
0.04 dB
±
0.02 dB
±
0.06 dB
±
0.04 dB
±
0.08 dB
Polarization Dependent Loss
(PDL)
±
0.025 dB
Scale Fidelity
<10 dB insertion loss
d
Auto Sensitivity
Full Range
±
0.02 dB
±
0.05 dB
±
0.07 dB
Dynamic Range @ 7 pm offset
Manual Sensitivity
Auto Sensitivity On
65 dB
30 dB
Return Loss
30 dB
35 dB (typical)
Receiver Noise Level
–85 dBm
Max Safe Power
Total Safe Power
+10 dBm
a. Not including connector repeatability.
b. RSS of PDL, stability, and scale fidelity.
c. Assumes sufficient sensitivity used.
d. Insertion loss from the DUT, cabling, connectors, couplers, fixtures, and so on.