2-11
Function Reference
Backup/Restore Menu
Backup/Restore Menu
Accesses the analyzer utilities used to backup or restore internal memory.
Key Path
Save/Recall > Backup/Restore Menu
Bandwidth Marker Interpolation (Marker Setup Panel)
Turns the bandwidth marker interpolation on or off. When on, the bandwidth
markers will be placed at the exact number of dB (NDB) from the normal
marker if within the trace range. The position of the marker will be linearly
interpolated between the two nearest trace data points. The default state is
on. If interpolate is off, for negative number of dB (NDB) values, the band-
width markers will be at values closest to and more negative then the NDB
value. For positive NDB values, the bandwidth markers will be at values clos-
est to and more positive than the NDB values. This will typically result in a
wider bandwidth measurement.
Key Path
Markers > Marker Setup > Bandwidth Marker Interpolation
Related Functions
Active Marker, Delta Marker
Remote
Commands
CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:FUNCtion:BANDwidth
:INTerpolation
BW Marker Units (Marker Setup Panel)
Sets the bandwidth marker X-axis readout for frequency or wavelength when
the instrument is in a non-zero span. The available selections are nm,
µ
m, Ang,
GHz, and THz. The default unit is nm (nanometers). This setting controls all
four bandwidth markers. The normal and delta offset markers have their own
settings.
Key Path
Markers > Marker Setup > BW Marker Units
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com