4-18
Programming Commands
Measurement Instructions
Examples
:CONF:ARR:POW
:FETC:ARR:POW?
:READ:ARR:POW?
:MEAS:ARR:POW?
:CONF:SCAL:POW -10 dBm
:FETC:SCAL:POW? MAX
:READ:SCAL:POW? MIN
:MEAS:SCAL:POW? DEF
Query Response
The following line is an examp le of a returned string when
:MEAS:SCAL:POW? MAX is sent:
-5.88000
If six laser lines are located and :MEAS:ARR:POW? is sent, the follow-
ing string could be returned. Notice that the first returned number
indicates the number of laser- line values returned in the query. The
measurement units are in dBm.
6,-1.37001,-1.10001,-9.62000,
-7.94000,-7.01000,-1.04001