4-38
Programming Commands
CALCulate2 Subsystem
WLIMit:STARt:FREQuency
Sets the start frequency for the wavelength limit range.
Syntax
:CALCulate2:WLIMit:STARt:FREQuency {?|{<real> | MINimum | MAXimum }}
<
real
> is a frequency value that is within the following limits:
Constant
Description
MINimum
181.6924 THz (1650 nm)
MAXimum
wavelength limit stop value
Attribute
Summary
Non- sequential command
Preset
State: 181.6924 THz
*RST State: 181.6924 THz
SCPI Comp liance: instrument sp ecific
Description
This command sets the starting range for the wavelength limit in hertz.
The start frequency value must be less than or equal to the stop fre-
quency value or the start frequency will be clip p ed to the stop fre-
quency and a “Data out of range” error will be generated. The default
units for the <
real
> p arameter are hertz.
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that this command has
the time to complete before sending any more commands to the instrument.
Refer
to “Always force the Agilent 86120C to wait for non-sequential commands” on
page 3-12
for more information.
Summary of Contents for 86120C
Page 12: ......
Page 26: ...1 14 Getting Started Returning the Instrument for Service ...
Page 27: ...1 15 Getting Started Returning the Instrument for Service ...
Page 28: ......
Page 96: ...3 18 Programming Monitoring the Instrument ...
Page 128: ...3 50 Programming Lists of Commands ...
Page 236: ...4 108 Programming Commands UNIT Subsystem ...
Page 248: ......
Page 264: ...6 16 Specifications and Regulatory Information Product Overview ...
Page 269: ...7 5 Reference Menu Maps Appl s Menu ...
Page 271: ...7 7 Reference Menu Maps Display List by WL Menu Delta On Menu ...
Page 273: ...7 9 Reference Menu Maps System Print Menu ...
Page 274: ...7 10 Reference Menu Maps System Setup Menu ...
Page 284: ......
Page 292: ......
Page 293: ......