4-48
Programming Commands
CALCulate3 Subsystem
ASNR[:STATe]
Turns the average signal- to- noise ratio on or off.
Syntax
:CALCulate3:ASNR[:STATe] {?|{ ON | OFF | 1 | 0 }}
Attribute
Summary
Preset
State: off
*RST State: off
SCPI Comp liance: instrument sp ecific
Description
This command turns the average signal- to- noise calculation on or off.
Only one of the CALCulate3 calculations (ASNR, DELTa, DRIFt, or
SNR) can be turned on at a time. Turning on the ASNR calculation
while another calculation is on will generate a “Settings conflict” error.
When the calculation is first turned on, the lines measured in the cur-
rent measurement will be used as the reference values for the signal-
to- noise ratio. Subsequent measurements will average the noise values.
The signal values are not updated until the number of measurements
used to average the noise is greater than or equal to the COUNt value.
Note
Only one STATe command can be turned on at any one time. Attempting to turn
more than one state on at a time results in a “–221 Settings Conflict” error.
Refer
to “Measure delta, drift, and signal-to-noise” on page 3-14
for additional informa-
tion on selecting measurements.
Summary of Contents for 86120C
Page 12: ......
Page 26: ...1 14 Getting Started Returning the Instrument for Service ...
Page 27: ...1 15 Getting Started Returning the Instrument for Service ...
Page 28: ......
Page 96: ...3 18 Programming Monitoring the Instrument ...
Page 128: ...3 50 Programming Lists of Commands ...
Page 236: ...4 108 Programming Commands UNIT Subsystem ...
Page 248: ......
Page 264: ...6 16 Specifications and Regulatory Information Product Overview ...
Page 269: ...7 5 Reference Menu Maps Appl s Menu ...
Page 271: ...7 7 Reference Menu Maps Display List by WL Menu Delta On Menu ...
Page 273: ...7 9 Reference Menu Maps System Print Menu ...
Page 274: ...7 10 Reference Menu Maps System Setup Menu ...
Page 284: ......
Page 292: ......
Page 293: ......